43
7 Commissioning
7.2
Setting examples
7.2.1
Measuring the pH value with pH combination electrode
Layout
Electrical connection
See chapter 5 "Installation", page 13.
Task
Measuring range:
2 to 12 pH
Output signal:
4 to 20 mA
Temperature measurement Pt100
Control function:
Pulse width controller
Setpoint 1:
pH 6.5
Setpoint 2:
pH 8.5
pH measurement with automatic temperature compensation.
Data sheet
(1)
Transmitter/controller type 202551
202551
(2)
pH combination electrode on the main board
201020
(3)
Coaxial cable
202990
(4)
Two-wire shielded cable
202990
(5)
Compensation thermometer Pt100 on the main board
201085
(1)
(2)
(3)
(4)
(5)