7 Commissioning
50
Basic setting for optional board 1
Administrator level/Password/Parameter level/Optional inputs/Analog input 1
Basic setting for optional board 2
Administrator level/Password/Parameter level/Optional inputs/Analog input 2
Controller settings
See chapter 13.6.1 "Simple limit monitoring", page 94.
Operating mode
pH measurement
Signal type
4 to 20 mA
Start of scaling
-600 mV (depending on the two-wire
transmitter)
End of scaling
+600 mV (depending on the two-wire
transmitter)
Temperature compensation source
Optional input 2
Operating mode
Temperature
Signal type
Pt100
Connection type
2-wire
Summary of Contents for dTRANS pH 02
Page 66: ...9 Calibrating a redox measurement chain 66...
Page 89: ...89 12 Calibration logbook Next most recent successful calibration Briefly press the key...
Page 100: ...14 Setup program 100 Export data for processing in an external program...
Page 133: ......
Page 134: ......
Page 135: ......