11 Calibrating a sensor with a standard signal
80
11.4.2 Calibration of the temperature coefficient
Linear temperature coefficient
The transmitter is in "Measuring mode".
✱
Immerse the conductivity sensor in the sample medium.
Start the calibration, see "Ways to start the calibration", page 71.
✱
Select "LINEAR TEMP. COEF.".
The current sensor temperature flashes in the display (1).
✱
Enter the required working temperature and confirm your entry.
The LC display now shows the selected working temperature (flashing) (2).
This example is based on a conductivity sensor with a connected two-wire
transmitter.
(1)
The working temperature must be at least 5 °C above or below the reference
temperature (25.0 °C).
(2)