11 Calibrating a sensor with a standard signal
✱
Press the
key.
✱
Use the
and
keys to set the value to the actual conductivity.
✱
Press the
key; the relative cell constant determined by the device is
displayed (as a %).
✱
Use the
key to accept the relative cell constant or
the
key to reject the values.
The transmitter is in "measuring mode" and displays the compensated
conductivity of the solution.
Calibration is complete
After rinsing, the sensor can again be used to take measurements.