69
10 Calibrating an ammonia sensor
✱
With manual temperature entry, use the
and
keys to set the
solution temperature and confirm your entry with the
key.
✱
Wait until the display value has stabilized; then press
to continue
✱
Wait until the display value has stabilized; then press
to continue.
✱
Use the
key to accept the calibration result or
the
key to reject it.
The device returns to measuring mode.
Calibration is complete
After rinsing, the sensor can again be used to take measurements.
PGM
PGM
PGM
PGM
EXIT
IIf the following permissible limits of the calibration values are not observed in
the calibration procedure then an error is displayed at the end of the proce-
dure:
Zero point:
-312 ... 588 mV