7
2 General
2.2
Arrangement of the oxygen measuring system
Basic version, type 202610/80...
- Two-wire transmitter JUMO dTRANS O2 01
- 8 m attached cable
- Terminal/operating unit with button for calibration
The basic version is designed for direct connection to a PLC or a recorder. The
calibration function can be initiated locally. An optional setup program is
available for configuring the transmitter. The two-wire transmitter is powered
by a separate power supply (optional), e.g. JUMO TN-22, see Data Sheet
95.6024
PLC
Visualization
or similar
24 V DC supply
for the
JUMO dTRANS O2 01
two-wire transmitter
Temperature at meas. point
optional (Pt1000)
2-wire connection
signal output for oxygen
(4 — 20 mA)
Terminal box
(Basic version)
Two-wire transmitter for
dissolved oxygen (DO)
Summary of Contents for 202610.0