14 Appendix
14.1 Before configuration
If a number of instrument parameters have to be modified in the instrument,
then it is advisable to note them in the table below, and then modify these
parameters in the sequence given.
The following list shows the maximum number of parameters that
can be altered.
Depending on the type and configuration, your instrument may not
show some of the listed parameters.
Code words to unlock the individual levels,
Parameter
Selection / value range
Factory
setting
New
setting
see page
O
2
INPUT
O
2
UNIT
mg/l or %
mg/l
48
PRESSURE COMPENS.
HEIGHT AMSL or
ATMOSPH. PRESSURE
HEIGHT
AMSL
HEIGHT AMSL
0 — 3000 m
300
ATMOSPH. PRESSURE
500 — 1500 hPa
976
SALINITY
0 — 100%
0
PROTECT. MEMBRANE
YES or NO
NO
O
2
OFFSET
-5.0 — 5.0
0.0
FILTER TIME
0.0 — 100.0
3.0
SLOPE O
2
50.0 — 200.0%
100.0
TEMPERATURE INPUT
TEMPERATURE UNIT
°C or °F
°C
49
TEMPERATURE OFFSET
-5.0 — 5.0
0.0
FILTER TIME
0.0 — 100.0 s
1.0
T-COMPENSATION
AUTOMATIC or
CONSTANT
AUTO
MANUAL TEMPERATURE -10.0 to 60.0°C
18.0
ANALOG OUTPUT
SCALING 4 mA
-199.9 — 199.9
0.0
49
SCALING 20 mA
-199.9 — 199.9
20.0
ON OVERRANGE
3.6 or 22.0 mA
22.0
RELAY 1 or 2
LIMIT VALUE
-199.9 — 199.9
0.0
51
HYSTERESIS
0.01 — 99.9
0.01
PULL-IN DELAY
0.0 — 999.9
0.0
DROP-OUT DELAY
0.0 — 999.9
0.0