11 Operation
46
11.2 Display in measurement mode
(1)
Status display of configured contacts K1 and K2.
If the characters are visible, the corresponding contact is activated.
(2)
Indicates present alarm states (e.g. calibration alarm, measurement
out-of-range).
(3)
Signals that the setup cable is connected.
If DATA appears here, only the transmitter and the PC are
communicating. There is no communication between transmitter and
operating unit.
In the SETUP or DATA status, the measurements are not updated in the
LC display.
(4)
The instrument is in the service mode. The output current for dissolved
oxygen has reached the value that was previously configured and is
kept constant.
(5)
The measurement and the unit are shown here.
(6)
In this area, the measured temperature of the medium and the unit are
shown.
(7)
The present operational state of the transmitter (measurement or
calibration) is shown here.
K1 K2
ALARM
SETUP
SERVICE
Measurement 10.25
18.5°C
mg/l
(5)
(6)
(7)
(4)
(3)
(2)
(1)