13 Eliminating faults and malfunctions
Fault / malfunction
Possible cause
Elimination
Precautionary
measures
No reading
No sensor reaction
No supply
Check supply voltage.
30ff.
Check electrical
connection.
30ff.
Jumper is missing.
Check electrical
connection.
No flow
Check flow and
restore, if necessary.
“Technical data of the
two-wire transmitter”,
page 15.
Deposits on
membrane
Clean membrane.
Sensor damaged
Replace sensor.
Then recalibrate.
42ff.
Indicated value is too
high
Clean cell membrane
and / or:
wait for stable output
signal, then
recalibrate.
42ff.
Sensor damaged
Replace sensor.
Then recalibrate.
42ff.