
42
10 Calibration
10.1 General
When to
calibrate?
- during commissioning
- when the sensor module is replaced
- after lengthy breaks in operation without supply voltage
Application-
specific
calibration
cycles
- drinking water:
1 to 6 months
- monitoring of waterbodies
(river or lakes):
1 to 4 months
- municipal effluents:
2 weeks up to 3 months
- industrial effluents:
1 week to 2 months
10.2 Calibrating
General
The transmitter is matched to the specific characteristics of the sensor module
during calibration.
Since the measuring system of this transmitter does not require zero-point
calibration, calibration is performed as single-point calibration.
Calibration can generally be carried out in two ways:
- in vapor-saturated air, i.e. in close proximity to the water surface. The
sensor
must
, however, remain
dry
!
- in air-saturated water. For this, the water is aerated until it is saturated with
air.
Air calibration
As it is rather involved to produce air-saturated water and difficult to
reproduce, the simpler method of air calibration is recommended for industrial
measurements.
Preconditions
- Transmitter, terminal box / operating unit and indicator must be connected
and powered.
Chapter 7 “Installation”, page 30ff.
- The sensor module (membrane and/or the optional protective film) is
clean
and dry
on the outside.
- The transmitter is in the air, as close as possible to the water surface.
In many cases, it will be sufficient to clean the transmitter head,
Pressure fluctuations in the measured medium will affect the
output signal!
The sensor
must
remain
dry
during calibration!
Summary of Contents for 202610.0
Page 2: ......
Page 21: ...21 6 Assembly...
Page 23: ...23 6 Assembly...
Page 66: ......
Page 68: ......