6 Assembly
24
6.5
Float fittings
Application
Float fittings are used to accommodate the immersion fitting in which the
dTRANS O2 01 oxygen transmitter is installed. The fittings are primarily used
for measurement in open basins or waterbodies. Different immersion depths
are facilitated by various immersion lengths of the fitting.
The following points should be noted at the design stage:
- The fitting must be easily accessible, to allow regular maintenance and
cleaning of the transmitter or the fitting itself.
- When the depth of water fluctuates, it must be ensured that the fitting (and
with it the transmitter) does not strike the bottom of the basin or waterbody
when the water level is low.
- The suitability of the materials of the fitting and the transmitter (e.g.
chemical compatibility) must be checked by the system designer.
ø
80
1" pipe
60
140
40
150
Terminal head
PVC
Shackle
Two-wire transmitter
for dissolved oxygen