8 Setup program
8.1
Function
Configurable
parameters
The setup program, which is available as an option, can be used to adapt the
transmitter to specific requirements.
- Setting the measurement unit (mg/liter or %).
- Setting the mean atmospheric pressure or the height a.m.s.l.
- Setting the salinity (salt content)
- Setting the range limits
- Setting the response of the output to out-of-range
- Description of the measurement point (19-character TAG number).
- Setting the functions of the limit comparators (lk).
- Setting the digital filter.
- Setting or checking the slope of the sensor module.
- etc.
Connection
Data transmission from or to the transmitter can only take place
when it is connected to the supply
Chapter 7 “Installation”, page
30ff.
The setup interface and the analog output are electrically isolated.
When connecting the PC interface cable, it is absolutely essential
to ensure that either the supply of the transmitter or of the PC is
not
galvanically coupled to the earth (e.g. use a battery-powered
notebook).
If no power supply (supply isolator) is available, the two-wire
transmitter can be configured using a 9 V battery as a power
source.
Supply
for the
two-wire transmitter
JUMO dTRANS O2 01
- PLC
- Visualization
- or similar.
Plug-on adapter
(included in setup set)
2-wire connection
signal output for oxygen
(4 — 20 mA)
24 V DC
Two-wire transmitter for
dissolved oxygen
Interface cable
(95/00301315)