PAGE 17 of 47
BG0104 Rev 03.3
1.
Ensure the probe holder is in latched, upper position
(Fig. 49)
.
2.
Using the supplied 3 mm hex driver loosen the transverse adjustment screw
(Fig. 53)
and rotate the yoke about the vertical shaft achieving the desired
angle.
3.
Tighten the transverse adjustment screw
(Fig. 54)
.
To return the transverse adjustment to neutral
(90°)
. The probe holder must be in
the latched, upper position
(Fig. 49)
. Rotate the yoke until the stop post contacts
the base of the probe holder
(Fig. 55)
. Then tighten the transverse adjustment
screw.
4.6.4. Probe Holder Longitudinal Adjustment
To adjust the probe holder’s vertical angle for longitudinal scanning, follow these
steps:
1.
Ensure the probe holder is in latched, upper position
(Fig. 49)
.
2.
Using the supplied 3 mm hex driver
(Fig. 14)
, loosen the longitudinal
adjustment screw
(Fig. 56)
.
3.
Rotate the main body of the probe holder until it is at the desired angle
(Fig.
57)
.
4.
Tighten the longitudinal adjustment screw
(Fig. 57)
.
To return the longitudinal adjustment to neutral
(90°)
. Line up the longitudinal
adjustment indicator markers
(Fig. 58)
.
Fig. 56
- Loosen 3 mm screw
Fig. 57
- Rotate to position
Fig. 58
- Line up markers