P/N 960-100928R_Rev. 3 {EDP #192692}
© 2018, JAPAN CASH MACHINE CO., LTD.
6 - 1 5
Performance Tests
UBA-RC™ Series Banknote Recycler
Section 6
Performance Test without PC Procedure
Perform the following steps to execute UBA-RC
™
performance Tests No.2 through No.9 without a
PC:
1. Turn
the
UBA-RC
™
Power Switch to
OFF
.
2. Set the desired DIP Switches on the UBA
™
Trans-
port Unit and Recycler Unit to
ON
for the desired
Performance Test.
3. Turn
the
UBA-RC
™
Power Switch to
ON
. The
UBA
™
Transport Unit LED will flash at a
Green
and a
Red
Color rate when entering the Perfor-
mance Test Mode.
4. Set
the
UBA
™
Transport Unit’s 8-Position DIP
Switch No.8
OFF
in order to start a desired Test.
5. Set
the
UBA
™
Transport Unit 8-Position DIP
Switch No.8 back
ON
in order to end a Test. The
UBA-RC
™
will revert to Stand-by Mode Status
following each Test’s completion.
Return to Step 2 of this procedure if necessary to
perform another DIP Switch Performance Test.
B
ANKNOTE
A
CCEPTANCE
T
EST
Perform the following steps to execute the
Banknote Acceptance Test without a PC:
1. Turn
the
UBA-RC
™
Power Switch to
OFF
.
2. Set the desired DIP Switch Number on the UBA
™
Transport Unit’s and Recycler Unit’s 8-Position
DIP Switch to
ON
to select the desired Perfor-
mance Test (See Table 6-4).
3. Turn
the
UBA-RC
™
Power Switch
ON
. The
UBA
™
Transport Unit’s LEDs will alternately
flash at a
Green
and a
Red
Color Rate when enter-
ing the Performance Test Mode.
4. Set the 8-Position DIP Switch No.8 to
OFF
in
order to start the UBA-RC
™
Unit’s initialization
routine.
5. Check that the UBA
™
Transport Unit’s LEDs are
out (extinguished), and that the UBA-RC
™
Status
is in the Stand-by Mode.
6. Start the Test by inserting a Banknote.
7. Confirm the Banknote’s denomination by count-
ing the LED flashes between pauses
(Table 6-4 in this Section for flash count vs.
denomination value information).
This completes the Banknote Acceptance Test
without using a PC Procedure.
NOTE: The denomination value stored in
the RC1-Bin is set by DIP Switches No.1
to No.4 on the Recycler, and the denom-
ination value stored in the RC2-Bin is set
by DIP Switches No.5 to No.8 on the
Recycler. Refer to the specific Country’s
“Software Specification Sheet” Document
for more value setting details.