Operation
TM/
TMC/RM/RMC-2030GE
- 61 -
10.5.6 PWC PIV Mode
The PWC PIV is based on PIV Fixed Exposure. In this mode, the first time exposure is controlled by the
pulse width of the external trigger, which is similar to PWC mode. The real exposure time of the first
image is equal to the pulse width of the external trigger.
The maximum trigger frequency in this mode is equal to 1/ (transfer time of two exposure
time of the first image).
The minimum active period of the external trigger is 10 pixel clocks (250ns)
Figure 60. PWC PIV Timing Table.
Table 8. Asynchronous Mode Chart
Asyn No Shutter
Async Preset and Prog. Shutter
PWC
aA
<1 line
<1 line
6 clk
aB
9.5 line
(n+1) lines + 1024 clk
Pulse width + 1024 clk
aC
1024 clk
Asyn No Shutter
Async Preset and Prog. Shutter
PW C
aA
<1 line
<1 line
6 clk
aB
9.5 line
(n+1) lines + 298 clk
Pulse width + 48 clk
aC
48 clk
PIV Fixe Expo
PIV PW C
Unit
pA
6
6
pB
70
70
pC
200
200
Pixel
pD
320
320
pE
20
20
pF
1
1
Frame