Editing Milling Jobs
84
NOTICE! Clean the filter, release the coolant fluid into the water circulation.
Damage to the device.
Danger to the environment.
Do not rinse the filter, but instead, change and dispose of as described in this chapter.
Dispose of the tank contents according to the specifications (see safety data sheet of the
coolant concentrate PrograMill Fluid for One, “Supporting documents”, page 9).
NOTICE! Operation of the unit without the addition of the coolant fluid.
Increased wear.
Reduced service life of the device and filter.
Use distilled water (according to VDE0510).
Always add the coolant concentrate PrograMill Fluid for One to the water.
NOTICE! Falling or tipping of the tank.
Damage to the tank.
Carefully remove the tank, fill and reinsert.
Ensure that the tank can not fall or tip.
7.2.1 Preparing the coolant fluid / filter change
Press in the indentation (1).
For easier removal, the tank is pushed
out a little.
You can now change / refill the water of the tank (“Changing / Refilling the water of the
tank”, page 85).
Or: You can now change the filter (“Finalizing coolant fluid / filter change”, page 88).