923 milliK manual - issue: 1.10
to
select channel 1 and then use the
↑
↑
↑
↑
↓
↓
↓
↓
keys to select each parameter in turn and press the
↵
↵
↵
↵
key to enter their values:
Press the
OK
(F4) key to start the calibration adjustment. Once started, the calibration
adjustment process cannot be aborted.
5.4 Exporting Calibration information
A calibration file containing the complete calibration history of the milliK in comma
separated (CSV) form can be exported to a USB drive for transfer to a PC. The file can be
imported into Excel™ or other suitable applications for viewing. To export the file connect a
USB drive to the USB port on the rear of the milliK, go to the ‘Calibration Menu’ Window
and press the
Export
(F3) key:
For a milliK with serial number 11-P123, the file will be stored to the USB drive at
:\milliK\11-P123\Calibration\CalibrationFile.txt