
© Isothermal Technology
Page 66 of 102
923 milliK manual - issue: 1.10
5.3 Adjusting the Calibration
It is common (good) practice to check the calibration of an instrument before making any
adjustment (see section 5.2 for information on checking the calibration of your milliK). After
adjusting the calibration of your milliK, its calibration should again be checked and the
values used to generate a report on its post-adjustment calibration.
The frequency of calibration is for the user to determine, but Isotech recommend that you
calibrate your milliK on an annual basis (the specification for the milliK is given in terms of
an initial calibration and additional uncertainty over a 1 year period). The values used to
adjust the calibration, together with information on the standards used and the calibration
conditions are all stored in a comma-separated (CSV) file that can be exported for viewing on
a PC, giving the whole calibration history of the instrument from its initial factory calibration.
The milliK includes calibration software that allows you to adjust all the calibration
parameters. Adjustment simply requires you to attach reference standards (calibrated
resistors, voltage source or current source) to the milliK and invoke the calibration features.
Typically a calibration adjustment involves the following:
Adjust 100
Ω
reading on the 115
Ω
range
Adjust 400
Ω
reading on the 460
Ω
Adjust 500k
Ω
reading on the 500
Ω
Adjust 0mV reading on channel 1
Adjust 0mV reading on channel 2
Adjusting 100mV reading on the 115mV range
Adjusting 10mA reading on the 30mA range
Adjusting the offset for the reference junction sensor
5.3.1 Starting the Calibration Adjustment Feature
Press the
Instrument
(F4) key, use the
↑
↑
↑
↑
↓
↓
↓
↓
keys to select ‘Calibration’ and then press the
↵
↵
↵
↵
key to start the calibration. You will be asked to enter the calibration password (see section
3.11) after which the ‘Calibration’ Window will appear, use the
↑
↑
↑
↑
↓
↓
↓
↓
keys to select each
parameter in turn and then press the
↵
↵
↵
↵
key to enter information that allows the calibration to
be identified. This information is stored in the calibration file for information purposes only
and you will not be allowed to proceed without entering something (at least one character) for
every parameter: