© Isothermal Technology
Page 37 of 102
923 milliK manual - issue: 1.10
Channel 1
Channel 2
Channel 3
Instrument Name
milliK
milliK
milliK
Serial Number
10-P196
10-P196
10-P196
Firmware Version
1.06
1.06
1.06
Software Version
1.3.5
1.3.5
1.3.5
Enabled
Yes
No
Yes
Range
115Ω
460Ω
30mA
Units
Ω
°C
°C
Current/mA
1mA
1mA
n/a
Reference Junction
n/a
n/a
n/a
Samples per Reading
1
1
1
Readings in Stats
100
100
100
Thermometer
Default 4-Wire
PRT
Default 4-Wire
PRT
0-200C 4-20 TX
Type
PRT
PRT
4-20mA
Manufacturer
Model
Default 4-Wire
PRT
Default 4-Wire
PRT
Serial Number
Recalibration Date
31/12/2099
31/12/2099
31/12/2099
Minimum Temperature
-273.15
-273.15
0
Maximum Temperature
2000
2000
200
Temperature Conversion
None
IEC60751 (2008)
Linear
Start Date
15/05/2012
Start Time
12:51:56
Elapsed Time/s
Channel 1
Channel 2
Channel 3
4246.2
100.001157
75.0036
4247.02
100.001152
75.0036
3.8.3 Clearing Logged Data from Internal Memory
The internal memory has a capacity of almost 4GB, which is sufficient to allow continuous
logging at the maximum rate for over 30,000 hours (1,250 days). Clear logged data from the
internal memory (to free capacity or discard data) by pressing the
Instrument
(F4) key and
using the
↑
↑
↑
↑
↓
↓
↓
↓
keys to select
Clear Logged Data
: