923 milliK manual - issue: 1.10
Press the
OK
(F4) key to start the calibration adjustment. The milliK will then take a number
of readings of the attached resistance standard, calculate and then apply an appropriate
adjustment. A progress bar appears showing the status of the calibration adjustment:
Once started, the calibration adjustment process cannot be aborted.
5.3.3 Adjusting PRT (460
Ω
) Resistance Range
Connect a calibrated (nominally) 400
Ω
resistor to channel 1. In the ‘Calibration Menu’
Window, use the
↑
↑
↑
↑
↓
↓
↓
↓
keys to select ‘460
Ω
Range’ and then press the
↵
↵
↵
↵
key to proceed to
the ‘460
Ω
Calibration Adjustment’ Window and use the same procedure described for
calibrating the 115
Ω
range (see section 5.3.2) to adjust this range.
5.3.4 Adjusting Thermistor (500k
Ω
) Resistance Range
Connect a calibrated (nominally) 500k
Ω
resistor to channel 1. In the ‘Calibration Menu’
Window, use the
↑
↑
↑
↑
↓
↓
↓
↓
keys to select ‘500k
Ω
Range’ and then press the
↵
↵
↵
↵
key to proceed to