923 milliK manual - issue: 1.10
Press the
OK
(F4) key to start the calibration adjustment. Once started, the calibration
adjustment process cannot be aborted.
5.3.8 Adjusting the RJC Sensor Calibration
Before you can use the RJC calibration adjustment feature you must first attach a calibrated
thermocouple to one of the channels and configure the channel for use with this thermometer
(see section 3.3.4). If you measure the ambient temperature with a calibrated PRT and use
this as the “known” temperature for the calibration (see section 5.2.6 for details), you can
reasonably use a good quality, uncalibrated thermocouple (the difference between ambient
temperature and the reference junction will be very small, so any error will not be
significant). If you use an uncalibrated thermocouple, make sure that you set the
thermocouple ‘Conversion’ to the thermocouple type used:
From the ‘Calibration Menu’ Window, use the
↑
↑
↑
↑
↓
↓
↓
↓
keys to select ‘RJ Sensor’ and then press
the
↵
↵
↵
↵
key to proceed to the ‘RJ Sensor Calibration Adjustment’ Window. Use the
←
←
←
←
→
→
→
→
keys