Adjustment Mode
69
4. Remove the weight. Display reads: Span= 20000, Weight= 0.000.
5. Repeat from step 1 if necessary.
6. On the left side of the machine, remove the bottom hex screw covering the
Span
button.
7. Press the
Span
button with a small screw driver to save the calibration parameters. A long beep will sound.
8. Press the
Mode
key then the
Enter
key to exit.
9. Verify the weight readings in the operation mode.
10. Seal the scale according to local regulations.
4.9
C10-Download
Update the scale’s firmware via a USB flash drive. Insert the USB
flash drive containing the soft folder before entering step C10.
USB specifications: 8 GB or smaller, FAT32 file system and USB 3.0 high-speed format.
The USB flash drive must be empty except for the soft folder.
Display
Description
Choices
C10-01 Copy Method
Selects the downloading procedure.
Enter the number and press
Enter
.
Select step 0 or 1 to download firmware
to the scale.
0: Firmware and default images
1: Firmware only
After pressing the
Enter
key, press the
Down arrow
key to display the firmware
version.
0. USB>MAIN (PRG+IMG)
1: USB>MAIN (ONLY PRG)
2. USB>MAIN (ONLY IMG)
3. USB>MAIN (BOOT REN)
4. MAIN>USB (PRG+IMG)
5. MAIN (ALL CLR)
3 and 5 are only available when booting
from a USB flash drive.
C10-02 USB Memory Folder
The firmware version is displayed. Press
the
Enter
key two times to begin
loading. When loading is complete
(message: Reboot Check 15029-0000)
power off the scale and remove the
USB flash drive.
--
Table 4-7. C10 Download
Hex Screw Covering
Span Button
C1O DOWNLOAD
&'2:1/2$'
Note
Important
Note