Invacare® AVIVA™ RX series
Problem
Other symptoms
Possible cause
Solution
Documentation
Motors stop and start
again
None
Voltage decline
Stop driving and allow
electronics to cool
down.
Motor runs but loses
power
None
High motor load
allows power module
to lower voltage
Stop driving and allow
electronics to cool
down.
None
High motor load
allows power module
to lower voltage
Leave mobility device
switched on and
let power module
operate. Charge
batteries overnight
with electronics
turned on.
None
Circuit breaker
switched off
Check cabling and
switch on circuit
breaker.
See
and
user manual.
None
Motor defective
Measure internal
resistance of motor,
replace motor if
defective.
See
and
.
Motors stop and do
not start again
None
Power module
defective
Replace power
module
See
.
Motors lose power
while driving
None
Bad connections
Turn off electronics,
wait 10 seconds, turn
on electronics. Check
all cabling.
See
Clutch(es) defective
Replace clutch.
See
.
Bearing defective
Replace motor.
See
.
Motor judders or runs
irregularly, or only
one motor runs
None
Collector defective
Measure internal
resistance of motor,
replace motor if
defective.
See
and
.
Bad connections
Check all cabling.
See
Circuit breaker
switched off
Check cabling and
switch on circuit
breaker.
See
and
user manual.
Batteries defective
Replace batteries.
See
.
Cabling to power
module or remote
defective
Check cabling.
See
None
Power module
defective
Replace power
module.
See
.
Motors do not run
Corroded contacts
Contacts defective
Check cabling, replace
if necessary.
See
20
1654647-B