6.11.2 Replacing Bus Expansion Block (GLM-CONX4)
The bus expansion block allows you to increase the amount of available bus sockets.
•
Phillips screwdriver, size 2
1.
Unplug all bus cables.
2.
Loosen and remove screws
B
.
3.
Replace block
A
.
4.
Install parts in reverse order. Tighten screws to 4 Nm.
5.
Plug in all bus cables.
6.
Test all functions.
Fig. 6-47
6.11.3 Replacing G-Trac Sensor
•
4 mm Allen key
•
10 mm wrench
Fig. 6-48
Removing sensor
1.
Remove batteries. See
2.
Unplug sensor cable from power module.
3.
Loosen and remove nut
E
and washer
D
.
4.
Remove bolt
B
and sensor
C
.
Installing sensor
CAUTION!
Risk of injury and damage due to uncontrolled movement of mobility device
An incorrect installed sensor sends wrong data to the power module.
– Ensure that sensor is installed with cable pointing sidewards.
– Ensure that nib on backside of sensor is engaged in its installation hole.
1.
Install parts in reverse order.
2.
Check that sensor is installed with cable pointing
sidewards and correctly engaged nib in its installation
hole
A
.
3.
Plug in sensor cable to power module.
4.
Test all functions.
6.11.4 Replacing Operating Hour Counter/ Connecting Cable