The Generator is intended for use in the electromagnetic environment specified below. The customer or user of the Generator should
ensure that it is used in such an environment.
Maximum Output
Power (Watts)
Separation (m)
150 kHz to 80 MHz
d = 1 .2√P
Separation (m)
80 to 800 MHz
d = 1 .2√P
Separation (m)
800 MHz to 2 .5 GHz
d = 2 .3√P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
Table 5—Recommended Separation Distances for the Generator, per IEC 60601-1-2, 3rd edition