EN - ENGLISH
11
INSTRUCTIONS FOR USE
Power On
Power on the Generator by pushing the power button. The power button will illuminate blue. An error in the start-up
indicates a device defect. Refer to the Error Diagnosis/ Troubleshooting section of this IFU.
Self-Test
The Generator performs a self-test after it is powered on. All display elements on the front panel are turned on during this process.
The following light up briefly in the order they are listed: The MONOCUT and BICUT mode buttons, then MONOCOAG and
BICOAG mode buttons, followed by the red indicator light above the neutral electrode receptacle. The two Output Power displays
light up with “88” and the white CUT and COAG Indicator lights illuminate sequentially; the respective activation tone must sound
simultaneously. The default start-up of the device is BICUT and BICOAG with both displays lit up with “00”.
An error in the self-test sequence indicates a device defect. Refer to the Error Diagnosis/ Troubleshooting section of this IFU.
Summary of Contents for Codman 901001ESUO
Page 2: ...2 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 137: ...EL 137 Codman Codman Codman Codman Codman Integra...
Page 138: ...EL 138 RF HF RF N2O...
Page 139: ...EL 139 4 5 3 2 3 0 10 C 50 F 10 C 50 F IEC 60601 1 16 70...
Page 142: ...EL 142 13 14 15 16 17 18 19 13 14 15 16 17 18 19 100 240 V 50 60 Hz T3 15 AH 250V...
Page 143: ...EL 143 20 21 22 23 20 21 22 CUT 23 COAG...
Page 145: ...EL 145 MONOCUT BICUT MONOCOAG BICOAG 88 CUT COAG BICUT BICOAG 00...