Guidelines and Manufacturer’s statement
HS112 C-1
ECG was manufactured for use within the electromagnetic environment stated below.
Buyer or user of the device
HS112 C
Test of disturbance
tolerance
IEC 60601
test level
Conducted RF
3 V
eff
IEC 61000-4-6
150 kHz – 80 MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz – 2,5 GHz
1
st
Note: In case of 80 MHz and 800 MHz, the higher frequency range must be applied.
2
nd
Note: This guideline must not be used for each case. Electromagnetic transmission is influenced by
absorption and reflection caused by buildings, objects and people.
a
It is not possible to define accurately the field intensity of stable transmitters, for example t
cellular phones, mobile radios, amateur radios, AM/FM radios and television broadcasting. In order to get to
know the electromagnetic environment of the stable R/F transmitters, on site measurements shall be made. In
case the field intensity at the site of the operation of
the appropriate operation of the device must be checked. In case improper operation is detected, further
arrangements shall be taken, for example it may be necessary to r
at a different site.
b
Field intensity above the frequency range of 150 kHz and 80 MHz must be lower than [V1] V/m .
HS112 C-1 User Manual
Guidelines and Manufacturer’s statement – electromagnetic disturbance
tolerance
ECG was manufactured for use within the electromagnetic environment stated below.
HS112 C-1
must ensure that the unit is operated within this environment.
IEC 60601
test level
Conformity
level
Electromagnetic environment
[V
1
] V
Portable and mobile RF telecommunication
80 MHz
devices must not be used within defence
3Veff
distance - defined by the
transmitter- from
HS112 C
2,5 GHz
[E
1
] V/m
3V/m
Suggested defence distance
d=[3,5/V
1
]√P
d=[3,5/E
1
]√P 80 MHz
d=[7/E
1
]√P 800 MHz –
P= maximum output in Watt (W) as per
manufacturer of the transmitter
d=suggested defence distance in meter (m)
Field intensity
a
of the stable RF transmitters
measured by tests on the spot must be smaller
than conformity levels
frequency range.
Disturbance may occur in the environment of
devices marked by the following sign :
Note: In case of 80 MHz and 800 MHz, the higher frequency range must be applied.
guideline must not be used for each case. Electromagnetic transmission is influenced by
absorption and reflection caused by buildings, objects and people.
It is not possible to define accurately the field intensity of stable transmitters, for example t
cellular phones, mobile radios, amateur radios, AM/FM radios and television broadcasting. In order to get to
know the electromagnetic environment of the stable R/F transmitters, on site measurements shall be made. In
y at the site of the operation of
HS112 C-1
exceeds the relevant RF conformity level,
the appropriate operation of the device must be checked. In case improper operation is detected, further
arrangements shall be taken, for example it may be necessary to rotate the device or to install the equipment
Field intensity above the frequency range of 150 kHz and 80 MHz must be lower than [V1] V/m .
R-6730-EN02-E
electromagnetic disturbance
ECG was manufactured for use within the electromagnetic environment stated below.
must ensure that the unit is operated within this environment.
Electromagnetic environment – guidelines
Portable and mobile RF telecommunication
devices must not be used within defence
defined by the frequency of the
HS112 C-1
and its cabling.
Suggested defence distance
√P 80 MHz – 800 MHz
– 2,5 GHz
maximum output in Watt (W) as per
manufacturer of the transmitter
d=suggested defence distance in meter (m)
of the stable RF transmitters
measured by tests on the spot must be smaller
levels
b
related to certain
Disturbance may occur in the environment of
devices marked by the following sign :
guideline must not be used for each case. Electromagnetic transmission is influenced by
It is not possible to define accurately the field intensity of stable transmitters, for example transmitters of
cellular phones, mobile radios, amateur radios, AM/FM radios and television broadcasting. In order to get to
know the electromagnetic environment of the stable R/F transmitters, on site measurements shall be made. In
exceeds the relevant RF conformity level,
the appropriate operation of the device must be checked. In case improper operation is detected, further
otate the device or to install the equipment
Field intensity above the frequency range of 150 kHz and 80 MHz must be lower than [V1] V/m .