HS112 C-1 User Manual
R-6730-EN02-E
57/73
14.
Device warnings
The possible warnings and recommended actions are listed below.
Memory OK
No action required.
Memory error
Due to an error, the stored data was damaged, and
the device has reset to factory settings. No action
required.
Not enough memory
Occurring with the Diagnosis or Store function; an
older record has to be deleted.
The heart frequency
is too high
A diagnosis can be obtained only under a 175 BPM
frequency.
There is no stored
record yet
Appearing on entering the catalogue in case there are
no stored records.
The printer is
overheated
Printing can start in approximately 10 minutes.
Printer error
Make sure that the paper in the tray is loaded
properly, and that the tray is closed.
The battery is
depleted
The battery has to be charged, or the device has to be
operated by connecting it to the mains power supply.
Check the electrodes
Appearing when the device detects an electrode
attachment error while printing or starting a
program. Check the leads in the 12-channel mode.
The ECG is too
noisy
Check the electrodes, turn the filters on.
There are few
recognizable beats
Appearing during diagnosis creation if the number of
beats is too low.
15.
Computer connection
The
Innobase for Windows
cardiological database management program can be purchased
for the
HS112 C-1
device. The program manages the patient database, stored records, it
performs 12-channel monitoring, record display and evaluation, and it can print in an A4
format. Using the
Innobase
program is described in a separate manual.
The device can be connected to a computer via a USB cable. The required minimal PC
configuration could be find in the Innobase user manual.
16.
Starting the demo mode
The device has a pre-programmed demo mode that can be used when the device is being
exhibited, or for presentational purposes. In order to run the demo mode, touch the screen
with 3 fingers immediately after turning the device on (at most 3 seconds after an image is
displayed on the screen). Exiting the demo mode can be done by restarting the device.