HS112 C-1 User Manual
14/73
R-6730-EN02-E
In order to do this, discharge the battery completely by disabling the automatic shutdown feature of the device
(
Setup/Others/Turn off
), then recharge it fully by connecting it to the mains power supply for at least 3 hours.
If battery capacity has decreased considerably (the battery discharges rapidly), it has to
be replaced. Battery replacement should be done by a qualified expert with a battery
pack acquired from Innomed Medical. If it is maintained in accordance with the
instructions given in the user manual, the minimal lifespan of the battery is 2 years.
Refer to the maintenance manual for details on battery and power failure.
Factory reset
In case of malfunction of the device you can reset the factory setting by switch on the device
with more than 10sec pressing of the ON button. Press it while you hear a sound. If during
pressing the GREEN LED become off, release the button, and immediately press it again for
more than 10 sec (until appearing the sound).
Caution!
The factory reset clear all memory, all records, and set the factory default setting.
5.2. Printer setup, loading the paper
Before first use, or if the device has run out of paper, a new stack of recording paper has to be
placed into the paper tray. Loading the paper tray consists of the following steps:
1.
Press the button indicated by the arrow, and pull out the paper tray.
2.
Place a stack of paper into the tray, and fold the top sheet over as indicated in the
figure below. The black page boundary markers and the raster grid should be facing
upwards.