5 - 10
Transpector MPS Operating Manual
5.5 Using Transpector Web UI
This section provides instruction for common Transpector MPS applications.
5.5.1 Transpector Web UI and FabGuard Explorer
CAUTION
FabGuard and Transpector Web UI cannot access the
same sensor at the same time.
FabGuard Explorer is the primary software package for the Transpector MPS
sensor, but there are instances when using Transpector Web UI is appropriate.
FabGuard Explorer will give control of the Transpector MPS sensor to the Web UI
if the RGA is idle.
5.5.2 Standalone Transpector Web UI Operation
This section provides instruction for using Transpector Web UI as a standalone
software program.
5.5.2.1 Data Acquisition
The
MONITOR
window is used for data acquisition.
NOTE:
Transpector Web UI does not allow for saving data or displaying the data
in a Trend graph.
Transpector Web UI is designed for data monitoring and
simple vacuum diagnostics.
When the
MONITOR
window is first opened, Transpector MPS will begin collecting
data using default parameters shown in
Table 5-2 Default Monitor Parameters
Parameter
Value
Min (AMU)
0
Max (AMU)
50
Dwell
32 ms
ppAMU
10