6 - 8
Transpector MPS Operating Manual
4
Carefully remove the new filament assembly from its packaging and place it on
the ion source standoffs.
CAUTION
The new filament assembly must be carefully placed onto
the ion source. Excessive horizontal movement will
cause the damage to the anode.
5
Replace the three washers and locknuts and ensure that the anode (C) aligns
with the mating bore (D). See
.
Figure 6-5 Align anode to mating bore
6
Tighten the washers (A).
7
Insert the filament leads back into the clamps and tighten.
8
Using an ohmmeter, check that the filament is not shorted to ground and that
the filament shows a resistance of approximately 0.8
at the feedthrough.