Transpector MPS Operating Manual
NOTE:
ALARM/WARNING SETTINGS and Transpector MPS Scan Parameters
each have a
SET
button next to the changeable parameters.
The
SET
buttons are not interchangeable.
To change alarm and warning settings, click
SET
next to the
ALARM
and
WARNING
group.
To change scan parameters, click
SET
next to the
MASS
and
DWELL
group.
3
Thermometer Bar—
The Thermometer bar shows the real-time level of the
signal. The bar will raise and lower based on the signal in amps of the specified
mass. If the signal exceeds warning or alarm levels, the color of the bar will
change. See
.
Figure 5-13 Thermometer bar levels
4
EMISSION and EM Control—
These work the same as
EMISSION
and
EM
in
the
MONITOR
window (
section 5.5.2.1.1, Editable Parameters, on page 5-11
).
Click
EMISSION
or
EM
to turn on or off the emission or EM.
5
Mass Trend Graph—
The mass trend graph will display how the signal
(in amps) for the specified mass changes with time. There are two lines
(one yellow and one red) on the trend graph. These lines represent warning
and alarm levels. When the signal increases above the alarm level lines, there
is the presence of a leak.