77
Mobil-O-Graph
®
, operating instructions, Rev. 5.1 10/2015, GmbH
6. Troubleshooting
6.1 Principal sources of error
The following may cause faulty measurements or undesired results:
■
Arm movement by the patient during the measurement
■
Switching off the device (e.g. at night)
■
Incorrect cuff size fitted
■
Slipping of the cuff in use
■
Absence of successful manual measurement in practice
■
Failure to take medication
■
Incorrect protocol set by the user
■
Batteries not fully or correctly charged, or worn out
■
Defect in the charging unit
6.2 Battery voltage curve
Figure 6.2.1 Rechargeable batteries
OK / Batteries OK.
Figure 6.2.2 Batteries old or too little
voltage, wrong batteries
6.3 Transmission errors
To avoid errors in data transfer, the monitor checks the data transmitted. If an error has occurred, E004 appears
on the display.
6.4 Check list
Please run through the following check list if errors occur when using the Mobil-O-Graph
®
. Many faults are eas-
ily remedied.
■
Check whether all cables are securely connected, and the infrared interface is correctly pointed towards
the monitor.
■
Check whether the monitor, computer and printer are switched on (depending on version obtained).
■
Check whether the connector cable is fitted to the correct serial interface (COM1 to COM4).
■
Check whether the batteries are adequately charged.
Note:
Some errors are associated with a continuous alarm for safety reasons. The continuous alarm can be
stopped by pressing any key. If there is pressure remaining in the cuff, open the cuff immediately.
6.5 Error tables
6.5.1
Troubleshooting, Mobil-O-Graph
®
Error symptom
Possible cause
Remedy
The time and date are not set to an
updated status after a longer peri-
od of time without power supply
through accumulators or batteries
The internal buffer battery is empty.
The biennial calibration has not
been performed.
The buffer battery is changed at
the metrological check.
The date and time can be set after
each change of batteries.
Send the device for calibration
to your dealer or directly to IEM
GmbH.
The measurement data can no
longer be called up or displayed.
An error has occurred during sav-
ing of patient data.
Delete the patient concerned
(menu bar) and re-enter.