40
Mobil-O-Graph
®
, operating instructions, Rev. 5.1 10/2015, GmbH
4.
H
ypertension
M
anagement
S
oftware Client-Server
►
In the toolbar, click
.
The window
Prepare monitor
appears. Its content depends on the type of monitor.
■
Mobil-O-Graph
®
■
Stabil-O-Graph
®
(optional)
4.8.1
Setting the measurement protocol
Note:
only possible for Mobil-O-Graph
®
Here you arrange how the 24-hour monitoring will be performed. There are eleven different protocols. Protocols
1 and 2 are user-definable..
Protocol 10 transmits each value automatically after measurement to your doctor’s PC via the Bluetooth inter-
face, or to a database via mobile communications such as a mobile phone or modem. You can use Protocol 10
for monitoring in practice, or for telemonitoring.
For practice monitoring you will need a Bluetooth interface on your PC. Steps 1-6 under 4.7.2 must be performed
and the patient in question opened in the HMS CS.
If you are interested in telemonitoring or require further information regarding practice monitoring, you can obtain
this from your specialist dealer or directly from IEM GmbH.
Protocol 11 activates the additional pulse wave analysis. The intervals for blood pressure monitoring and pulse
wave analysis can be configured independently. A PWA will always precede measurement of blood pressure,
however.
Selecting a protocol
►
In the pop-up field
Protocol
click the protocol you require.
►
In Protocol 1 or 2, set the timeframe (beginning of time interval) under
daytime interval
and
night interval
,
■
timeframe (beginning of time interval),
■
the number of measurements during this interval,
■
whether the measurement will be shown on the monitor (Display measurements),
■
and whether an acoustic signal (buzzer) will sound as the measurement starts.
Transmit protocol
►
Click
.
►
Confirm with
4.8.1