4.
H
ypertension
M
anagement
S
oftware Client-Server
35
Mobil-O-Graph
®
, operating instructions, Rev. 5.1 10/2015, GmbH
Begin 24-hour monitoring
Starting situation:
■
Mobil-O-Graph
®
is switched on.
Step
Mobil-O-Graph
®
PC
1)
►
Put the monitor on the patient (put on the cuff
and connect it to the recorder).
►
Press
for a manual measurement, to
ensure that the monitor works as it should.
►
Wait for the first, manual measurement.
If everything is in order, the patient can leave.
A successful measurement is necessary to
activate the protocol!
Read out measurements from 24-hour monitoring from Mobil-O-Graph
®
via Bluetooth
Starting situation:
■
Mobil-O-Graph
®
is switched on,
■
PC is switched on,
■
Mobil-O-Graph
®
interface recognises HMS CS.
Step
Mobil-O-Graph
®
PC
1)
►
Start HMS CS.
The active Bluetooth connection is
shown in the Toolbar.
The window
Action of device
appears.
2)
►
Take the monitor off the patient (take off the
cuff and disconnect the recorder)
►
Hold down
nd press
.
bt
flashes on the display.
►
Press
.
bt
stops flashing and the buzzer sounds.
3)
Buzzer sounds twice.
4)
►
Click
.
The question appears:
Assign measurements to
patient with ID number xxx?
►
Click
.
After data transmission the window
Measurement
series appears.
►
If necessary, change the times for the daytime
and night intervals.
►
Click
.
The question appears:
Assign measurements to
patient with ID number xxx?
►
Click
.
The transmitted measurements are displayed
as a Table of measurements.
4)
Buzzer sounds.
First
bt end
, and then the time appear on the dis-
play.
►
Switch off the Mobil-O-Graph
®
.
For further analysis of measurement series see 4.11,
Analysing measurements
.
4.7.2