36
Mobil-O-Graph
®
, operating instructions, Rev. 5.1 10/2015, GmbH
4.
H
ypertension
M
anagement
S
oftware Client-Server
4.7.3
Connection via infrared (IR)
The IR-Med infrared interface enables wireless data transmission between the Mobil-O-Graph
®
monitor and the
PC. The IR-Med infrared interface is connected to the PC by cable:
■
via serial interface (COM1,2...) or
■
via USB interface.
IR-Med infrared interface (right)
with serial interface (left)
Connecting the serial interface
to the PC
When using the Mobil-O-Graph
®
monitor with the
IR-Med
infrared interface, follow these steps:
■
Connect the monitor to the PC,
■
Configure the monitor in HMS CS
■
Set the monitor for the patient using HMS CS,
■
Test the monitor and put it on the patient.
After 24-hour monitoring
■
Take the monitor off the patient (take off the cuff and disconnect the recorder),
■
Read out measurements from monitor and save them in HMS CS under patient information.
Connect PC to Mobil-O-Graph
®
via IR-Med infrared interface
Step
Mobil-O-Graph
®
PC
1)
►
Switch off the Mobil-O-Graph
®
2)
Establish a connection between Mobil-O-Graph
®
and PC
a)
via serial interface (COM1,2...) with cable or combi cable
►
Switch off the PC
►
Connect the cable to the serial interface
(COM1,2...)
►
Switch on the PC.
b)
via USB interface
►
Connect the cable to a USB port.
3)
►
Point the Infrared interface of the Mobil-O-
Graph
®
towards the
IR-Med
infrared interface
at a distance of 10-15 cm.
4)
►
Switch on the Mobil-O-Graph
®
Change to co mode:
►
Hold down
and press
.
►
Press
until
co
flashes on the display.
►
Press
.
co
stops flashing and the buzzer sounds.
4.7.3