European User Manual
•
When the Extraction is done, you may try to extract the cryoprobe from the
tissue.
•
Pressing the
Finish
icon will terminate the procedure.
Figure 86: Extraction screen in Manual Mode
During the manual procedure, the system provides information regarding treatment
progress:
•
The left side of the screen displays the Current step (
Freeze
,
Thaw
or
Extraction
)
and time elapsed for this specific step.
•
The right side of the screen displays information on the entire Protocol as total
elapsed time and treatment format.
•
If you perform more than 3 consecutive
Freeze/Thaw
steps, the right side will only
display the last 5 actions due to screen size limitation (3
Freeze
-
Thaw
steps).
•
The system allows up to 7
Freeze/Thaw
steps
5.3.3.3
Automatic Freeze mode
Choosing one of the suggested protocols from the Treatment Selection screen will open
the automatic protocol screen.
•
To activate the
Freeze/Thaw
preprogrammed steps, press the Action button on
the straight cryohandle and hold it for one second or press the
Freeze
icon on the
left lower corner of the screen. The
BLUE
LED will light up around the action
button (optional).
•
The
Start
icon will change to a
Skip
icon.
•
The skip icon enables the user to skip from the current step in the cycle to the next
step as preprogrammed.