Figure 71:
Lesion size value
3.
Enter the Lesion size value in centimeters. After the Lesion size field is populated, the
planned margin options are presented at the bottom of the screen.
4.
Select the relevant option.
Figure 72:
Planned Margin
5.
Press
Calculate
to calculate the cryoprobe placement. The calculated result (X) that will
appear is the distance the Cryoprobe should pass through to ensure that the center of the
lesion is aligned with the center of the Cryoprobe cooling area.
If the entered lesion size value equals to or is larger than maximum (6 cm) or the
calculated cryoprobe placement value is less than -1 cm (Cryoprobe tip is likely inside
the lesion)
, “N/A” (not applicable) will appear in the calculation result field.