![IceCure ProSense FAS3000000 User Manual Download Page 108](http://html1.mh-extra.com/html/icecure/prosense-fas3000000/prosense-fas3000000_user-manual_610296108.webp)
IceCure™ Medical Ltd.
DMS-7064 rev. C
ProSense™
Confidential
User Manual
108
European User Manual
Figure 113: System Fail example screen
2.
If you are in the middle of a cryotherapy procedure,
wait for passive Thaw
and
then carefully
remove the cryoprobe
from the tissue.
Replace the cryoprobe
when required automatically by the system.
3.
Disengage the cryoprobe
from the cryohandle
only
after the system instructs you
that it is safe to do so.
4.
If system shut down is required,
shut OFF the
ProSense™
cryoablation system
by
pressing the mechanical ON/OFF switch on the main chassis and then remove the
plug from the power supply.
ProSense™
cryoablation system touch screen failure
If the touch screen shuts down, you will need to shut OFF the system with the mechanical
ON/OFF switch and remove the plug from the power supply. This will end the
cryoablation procedure.
Warning
In case of software crash, switch OFF the mechanical ON/OFF switch and
unplug the electrical cable.
Call
IceCure™
Medical for technical service before restarting the
ProSense™
cryoablation system.
5.5.2.1
Emergency Stop button
The fastest way to end a procedure is to press EXTRACT on the screen (or
the RED
HANDLE button).
However, in case of electrical emergency, press the RED ROUND
Emergency Stop button located on the right side of the main chassis. The Emergency Stop
button is designed for emergency shutdown of the unit. Pressing this button immediately
turns OFF the system.
Switch OFF the mechanical ON/OFF switch and unplug the electrical cable.