Problem
Probable Cause
Action
Unreasonable
temperature readings of
temperature sensor
Temperature sensor or system
malfunctions.
Check that the temperature sensor
is well connected.
If so, change the temperature
sensor.
If temperature reading is still
unreasonable, continue the
procedure without a temperature
sensor.
If problem persists, contact
IceCure™
Medical service.
Cryoprobe cannot easily
be removed from target
tissue.
EXTRACTION (WARM) process
malfunction
Do not use force in attempt to
extract the cryoprobe from the
tissue.
Wait a few minutes and try
removing it gently again.
If problem persists, contact
IceCure™
Medical service.
Cryoprobe is difficult to
disengage from
cryohandle.
Parts are frozen together
Try again after five minutes. If it still
cannot be disengaged, contact
IceCure™
Medical service.
Liquid nitrogen leak
when trying to remove
the cryoprobe
Excess pressure on the pipes, main
valve open, procedure not
completed, or Dewar relief valve
malfunction.
When initial leak is detected, screw
the cryoprobe back in place.
Make sure that the procedure has
ended, wait 3 minutes and try
removing the cryoprobe again. If
liquid nitrogen is still leaking contact
IceCure™
Medical service.
Computer failure due to
hardware fault
Unknown
Mechanical shut down of the system
using the ON/OFF switch at the back
of the system or emergency stop
button.
Then turn on the system (after
releasing the emergency stop
button, if pressed) and wait for
computer re-boot.
If this occurs during a procedure,
replace the cryoprobe when
required to do so by the system,
according to training materials
provided by I
ceCure™
Medical.
Screen is stuck/
Controller stops
working/ Controller is
stuck
Unknown
Mechanical shut down of the system
using the ON/OFF switch at the back
of the system or emergency stop.
Then turn ON the system (after
releasing the emergency stop
button, if pressed) and wait for
computer re-boot.
If there is a cryoprobe in the target
tissue, refer to training materials
provided by
IceCure™
Medical for
further details.