User Manual
118
European User Manual
TROUBLESHOOTING
General
Warning
Do not modify this equipment without authorization of the manufacturer.
Warning
Never open the console. Only an
IceCure™
Medical technician or authorized
representative is allowed to open the console for maintenance or to repair the
system.
Warning
Do not attempt to perform any troubleshooting or corrective action beyond
those specified in the following guide. Any malfunction not listed in the guide,
or one that persists after the recommended action has been taken, must be
referred to
IceCure™
Medical
.
Troubleshooting guide
Table 6: Troubleshooting guide
Problem
Probable Cause
Action
Main chassis does not
move
Rollers are locked
Unlock the rollers, transport the
system and lock them again.
If the problem persists, contact
IceCure™
Medical service.
No system power
AC power is not connected to the
system; power supply malfunction
Check that the power cable is
connected to the inlet and the wall
outlet.
Check that Emergency button is
released.
Check that main switch is on.
If still there is no power, contact
IceCure™
Medical service.
AC power is on but the
screen does not turn on
Computer power supply
malfunction or computer error
Check if the computer power cable
is connected to the screen and turn
the screen on.
If the screen still does not turn on
contact
IceCure™
Medical service.
The system and screen
turn on
but doesn’t
enter to the Main screen
AC power supply malfunction
Contact
IceCure™
Medical service