IceCure™ Medical Ltd.
DMS-7064 rev. C
ProSense™
Confidential
User Manual
44
European User Manual
Figure 25: Emergency Stop button
Caution
Use the Emergency Stop button only when there is no other choice.
Whenever possible, use
Extraction
to release the cryoprobe from the tissue
and use the standard shutdown procedure.
Excessive use of the Emergency Stop button may damage the system.
Warning
If a major electrical risk is detected during the cryotherapy procedure, use the
Emergency Stop button immediately.
Flexible hose
The hose joins the cryohandle to the siphon inside the chassis. This permits liquid
nitrogen to flow from the main chassis to the cryoprobe.
Caution
Do not pull the system by the cryohandle or hose.
The hose shouldn’t be bent or flexed
especially during procedure.
Touch screen and User Manual activation
The touch screen is located on top of the main chassis and allows for operating and
monitoring of the system. It is designed for users and technicians.
Do not connect any signal input/output port to the touch panel PC except certified
equipment provided by
IceCure™
Medical.
A sample touch screen display is represented by the figure below. For further information
about computer interface and operating the system, see Chapter 6.