Warning
Liquid nitrogen may cause
, if there is not sufficient ventilation in the room,
serious
injury or burn if handled improperly. Local laws and safety rules regarding liquid
nitrogen Dewars should always be observed. Maintenance of liquid nitrogen
Dewars should be performed by authorized personnel only.
•
Check that the Dewar is filled properly and cover it with its sponge lid to transfer it
back to the system.
•
Check there is no frost on the Dewar after you filled it. If there is a frost on the
Dewar opening, wipe gently the Dewar opening prior to insertion to the system.
•
In the event that a Dewar is damaged, use another Dewar to continue the
procedure.
Warning
To avoid spillage and cold burns, do not transfer a Dewar containing liquid
nitrogen unless it is covered with its designated lid.
Warning
Do not use a liquid nitrogen Dewar if it is damaged.
You can tell that a Dewar is damaged if after filling it, frost appears on the outer
wall of the container. Return the Dewar to an
IceCure™
Medical technician or an
authorized distributor for inspection.
Warning
At the beginning of the procedure the Dewar should be filled around 1 cm below
the neck of the bottle.
Do not overfill a Dewar with liquid nitrogen; it will reduce performance of the
system.
•
Take off the lid and place the Dewar in its position into the system; bottom first.
Make sure the Dewar moves freely (right-to-left) when you move the Dewar-
handle (see pictures below).