Dewar door is open
If the system detects that the Dewar compartment door is open when it tries to move the
Dewar carriage up or down, the following popup message will appear:
Dewar door is open. Close the door to continue.
Figure 59: Dewar door open warning
5.2.6.2
Sterile sleeves
After filling the Dewar and placing it in its compartment, please work under sterile
conditions. Cover all relevant accessories in the sterile area with the suitable single-use
sleeves, to ensure sterility.
5.2.6.3
Cryoprobe Selection
Select a cryoprobe according to your clinical judgment and target tissue.
The ice ball size is a function of the freezing time and is also affected by the tissue and
blood flow (see Figure 60).
On the label of the cryoprobe packaging, the specification of the
cryoprobe is clearly mentioned, with reference to the following
elements:
* Diameter (2.4 or 3.4mm)
* Shaft length (defined in mm)
* Ice ball shape (spherical or elliptic).
* Tip shape (trocar or pencil or blunt)