European User Manual
Caution
If one or more of the rollers is damaged, do not use the system.
Warning
Cryoprobes are fragile and can be damaged if mishandled. Do not use a
cryoprobe that has been bent, dropped, hit against a hard surface or
compromised in any manner, as damage to the cryoprobe may have
occurred.
Warning
1. While cryoprobe is in use, surgical instruments should be handled with
care to avoid damage to the cryoprobe.
2. Avoid pinching, cutting or pulling the cryoprobe. Do not use damaged
cryoprobes
3. Do not grasp cryoprobes with auxiliary instruments, as this may cause
damage to the cryoprobe shaft.
Warning
If a sterile product packaging is compromised in any way, do not use the
product.
Warning
Do not move the system when the Dewar contains liquid nitrogen.
Warning
Removing the Dewar or placing it back within the system after refilling it
must only be done according to system instructions and with the carriage in
the bottom position. If the carriage is not in the bottom position, liquid
nitrogen may spill out.
Warning
When replacing the Dewar, follow detailed on-screen instructions on
opening the Dewar compartment.