IceCure™ Medical Ltd.
DMS-7064 rev. C
ProSense™
Confidential
User Manual
20
European User Manual
Warning
Verify cryoprobe S/N registration by double checking the serial number on
the package and on the cryoprobe itself. Entering an incorrect cryoprobe S/N
registration will result in cryoprobe nullification or incomplete treatment.
Warning
Do not allow any liquid or humidity to enter the cryohandle. Always keep
the cover on the cryohandle.
Warning
In case the Extraction process isn't available, Wait for passive Thaw.
Warning
In case of frost on shaft, if possible, start
Extraction
. If not, wait for passive
Thaw
. In both cases use skin protection techniques.
Warning
Pay attention to the cryoprobe during functional test: bubbles, leak or frost
on shaft may signal internal problems that could be hazardous.
Warning
If you don’t press
Skip
after replacing Dewar, the system will stay in
Thaw
mode.
Liquid nitrogen
Nitrogen gas is a potential asphyxiant. In the event of a large liquid nitrogen spill,
personnel should adhere to a predetermined evacuation plan. Seek medical help
immediately if breathing problems occur.
Warning
If a liquid nitrogen leak is detected at any time, press the
Extraction
button
immediately.
Warning
At the beginning of the procedure the Dewar should be filled around 1 cm
below the neck of the bottle.
Do not overfill a Dewar with Liquid nitrogen; it will reduce performance of
the system.