8. Troubleshooting
8-3
i-Smart 300 Blood Gas Analyzer
Troubleshooting,
continued
Calibration
If the sensor state appears as slope error in red background,
1.
Run Cal 2.
2.
If needed, repeat additional Cal 2 a few times more.
Cartridge Data
Copy
If the analyzer fails to detect a USB memory,
1.
Remove the USB memory from the USB port.
2.
Insert the USB memory into a different USB port.
3.
If it still doesn’t work, try a different USB memory.
If “Cartridge data copy has failed” message appears during data
copy,
1.
Close the message and try again from the beginning.
2.
Select desired cartridge data and press
Copy.
Cartridge
Installation
If the analyzer does not change to the warming-up screen after the
cartridge is inserted into the analyzer,
1.
Check that the cartridge is an i-Smart 300 cartridge.
2.
Check that the cartridge is not past its expiration date.
3.
Check that the cartridge has not been previously used.
4.
Open and close the cartridge door.
Summary of Contents for i-Smart 300
Page 2: ...Operators should read the entire manual before installing and operating the analyzer...
Page 5: ...1 Introduction 1 Introduction Safety 1 1 Symbols and Terms 1 3 Product Description 1 5...
Page 12: ...1 Introduction 1 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 25: ...3 Analyzer Setting Operator s Manual 3 2 Screen Menus continued Menu options...
Page 48: ...3 Analyzer Setting 3 23 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 49: ...4 Sample Analysis 4 Sample Analysis Blood Samples 4 1 QC Samples 4 8...
Page 76: ...5 Database 5 13 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 77: ...6 Shutdown 6 Shutdown Analyzer Shutdown 6 1 Cartridge Restart 6 3...
Page 90: ...7 Maintenance 7 9 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 91: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 8 1 Error Code 8 6...
Page 98: ...8 Troubleshooting 8 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 104: ...9 Specification 9 5 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...