
2. Installation
2-1
i-Smart 300 VET Blood Gas Analyzer
Installation Requirements
Unpacking
Inspect the shipping package of the product for any signs of
damage from shipment.
Ensure all listed components are included in the package.
Environmental
Requirements
Location
:
indoors
Temperature
: 15 ~ 35°C
Humidity
: 5 ~ 85%,
relative humidity
Altitude
:
under 3,000 m
Note:
The analyzer should be installed on a horizontal, stable surface in
a well-ventilated area without direct sunlight or heat.
Allow enough space for the cartridge door of the analyzer to
open.
Allow at least 10 cm of space from the back of the analyzer for
adequate ventilation.
This equipment has been tested and found to comply with the
limits for a Class A digital device, pursuant to part 15 of the FCC
Rules. These limits are designed to provide reasonable protection
against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses, and
can radiate radio frequency energy and, if not installed and used
in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the
interference at his own expense.
Electrical
Requirements
Volt AC
: 100 ~ 240 Va.c.,
grounded electrical supply
Amp
: 1.5 A
Frequency
: 50/60 Hz
Power cord
:
3-wire, approved
Power adapter
: +24 Vd.c., 2.7 A (
Output
)
Summary of Contents for i-Smart 300
Page 2: ...Operators should read the entire manual before installing and operating the analyzer...
Page 5: ...1 Introduction 1 Introduction Safety 1 1 Symbols and Terms 1 3 Product Description 1 5...
Page 12: ...1 Introduction 1 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 25: ...3 Analyzer Setting Operator s Manual 3 2 Screen Menus continued Menu options...
Page 48: ...3 Analyzer Setting 3 23 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 49: ...4 Sample Analysis 4 Sample Analysis Blood Samples 4 1 QC Samples 4 8...
Page 76: ...5 Database 5 13 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 77: ...6 Shutdown 6 Shutdown Analyzer Shutdown 6 1 Cartridge Restart 6 3...
Page 90: ...7 Maintenance 7 9 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 91: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 8 1 Error Code 8 6...
Page 98: ...8 Troubleshooting 8 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 104: ...9 Specification 9 5 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...