6. Shutdown
6-3
i-Smart 300 Blood Gas Analyzer
Cartridge Restart
Power Recovery
When the analyzer recovers from power interruption, the
cartridge inside the analyzer can be continuously used, only if:
The analyzer was in Ready state or in process of calibration
when the power was interrupted and the power returns
within 1 hour.
The analyzer was analyzing a sample when the power was
interrupted and the power returns within 20 minutes.
The analyzer was analyzing a QC sample when the power
was interrupted and the power returns within 1 hour.
1.
When the analyzer recovers from power interruption with the
cartridge inside, one of following
Cartridge Restart
screens
will appear.
2.
The
Cartridge Restart
from power recovery will take about 10
minutes.
3.
After the
Cartridge Restart
, the analyzer will return to the
Ready state.
Summary of Contents for i-Smart 300
Page 2: ...Operators should read the entire manual before installing and operating the analyzer...
Page 5: ...1 Introduction 1 Introduction Safety 1 1 Symbols and Terms 1 3 Product Description 1 5...
Page 12: ...1 Introduction 1 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 25: ...3 Analyzer Setting Operator s Manual 3 2 Screen Menus continued Menu options...
Page 48: ...3 Analyzer Setting 3 23 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 49: ...4 Sample Analysis 4 Sample Analysis Blood Samples 4 1 QC Samples 4 8...
Page 76: ...5 Database 5 13 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 77: ...6 Shutdown 6 Shutdown Analyzer Shutdown 6 1 Cartridge Restart 6 3...
Page 90: ...7 Maintenance 7 9 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 91: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 8 1 Error Code 8 6...
Page 98: ...8 Troubleshooting 8 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 104: ...9 Specification 9 5 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...