3. Analyzer Setting
사용 설명서
3-12
Sample Setup,
continued
Critical Limits
1.
Go to
Menu > Setup
(
Setup
menu is protected by password.)
>
Critical Limits Setup
. The following screen will appear.
2.
Press each box and enter low and high values of the critical
limits range for each parameter.
Note:
To delete the entered data,
press
Current Tab Clear
.
Before saving the setup, the analyzer will validate the entered
values to confirm that:
-
Values for both low and high limits are entered.
-
Low limit value is lower than high limit value.
If any of the entered values are not valid, the analyzer will not
save the setup and ask to correct the invalid value at the setup
screen.
3.
Press
Save
to save the setup and go back to the main screen.
To exit to main screen without saving the changed setting, press
Close
.
Summary of Contents for i-Smart 300
Page 2: ...Operators should read the entire manual before installing and operating the analyzer...
Page 5: ...1 Introduction 1 Introduction Safety 1 1 Symbols and Terms 1 3 Product Description 1 5...
Page 12: ...1 Introduction 1 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 25: ...3 Analyzer Setting Operator s Manual 3 2 Screen Menus continued Menu options...
Page 48: ...3 Analyzer Setting 3 23 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 49: ...4 Sample Analysis 4 Sample Analysis Blood Samples 4 1 QC Samples 4 8...
Page 76: ...5 Database 5 13 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 77: ...6 Shutdown 6 Shutdown Analyzer Shutdown 6 1 Cartridge Restart 6 3...
Page 90: ...7 Maintenance 7 9 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 91: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 8 1 Error Code 8 6...
Page 98: ...8 Troubleshooting 8 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 104: ...9 Specification 9 5 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...